HeartSine - AED Samaritan PAD Technical Specifications
PHYSICAL Size: 8.0in x 7.25in x 1.9in (20cm x 18.4cm x 4.8cm) Weight: 1.1 kg (2.4 lbs) including Pad-Pak™ Battery.
DEFIBRILLATOR Waveform: SCOPE® (self-compensating output pulse envelope) Biphasic Escalating waveform. Optimized biphasic waveform compensates energy, slope and envelope for patient impedance. Energy Selection: Pre-configured automated protocol (escalating from 100J, 150J, to 200J). Charge Time: Less than 7 seconds (100J) and less than 12 seconds (200J) from a SA message with new Pad-Pak™. Shock Cycle Time: <60 seconds for three analyze/shock cycles. Controls: On/Off and Shock button
PATIENT ANALYSIS SYSTEM Method: Evaluates patient’s ECG, signal quality, electrode contact integrity and patient impedance to determine if defibrillation is required. Sensitivity / Specificity: Meets AAMI DF 80.
DISPLAY Visual Prompts: Attach Pads, Stand Clear, Perform CPR, Shock Now, Self Tests Pass - Ready State. Indicators: Low battery, (at least 10 discharges remaining), audible prompt (alerts user of electrode disconnect), service icon stops flashing if self-tests fail, no battery, or when service required. Voice and icon prompts: Extensive voice and icon prompts guide the user through the operation sequence. Languages: English
ENVIRONMENTAL Operating Temperature: 0° to 50° C (+32° to +112°F) Storage Temperature: -10° to 50° C Relative Humidity: 5% to 95% (non-condensing) Water Resistance: IEC 60529/EN 60529 IP44 Altitude: 0 to 15,000 feet (0 - 4,575 meters) Shock: MIL STD 810F Method 516.5, Procedure I (40G's) Vibration: MIL STD 810F Method 514.5 Category 4 Truck Transportation - US Highways Category 7 Aircraft - Jet (737 & General Aviation Exposure) EMC: EN 60601-1-2, Second Edition: 2001 Radiated Emissions: CISPR11: 1997 +1A:1999+A2:2001 Group 1 Class B Electrostatic Discharge Immunity: EN61000-4-2:1995 (8KV)+A1:1998+A2:2000 RF Immunity: EN61000-4-3:1996, 80 MHz - 2.5 GHz (10V/m)+A1:1998+A2:2000 Magnetic Field Immunity: EN61000-4-8:1993 (3 A/m)+A1:2000 Aircraft: RTCA / DO - 160D: 1997, Section 21 (Category M)
EVENT DOCUMENTATION Type: Internal memory Memory Capacity: 45 minutes of ECG (full disclosure) and event/incident recording. Playback Capabilities: Custom USB cable directly connected to PC and SAVER® windows-based data review software.
Pad-PAKTM BATTERY PAK Shelf Life: 3 years Weight: 0.44 lbs (0.2kg) Size: 3.93in x 5.24in x .94in (100mm x 133mm x 24mm) BATTERY Type: Lithium Manganese Dioxide (LiMnO2) 18V, 0.8 Amp Hrs Capacity: >30 shocks at 200J or 6 hours of continuous monitoring. Standby Life: 3 years ELECTRODES Electrodes: Samaritan® disposable defibrillation electrodes are supplied standard with each Pad-PakTM. Active Gel Area: 100 cm2 Cable Length: 3.5 ft (1m) Shelf Life: 3 years
 |